Abstract | A matrix-fibril shear stress transfer approach is devised and developed in this paper to analyse the primary biomechanical factors which initiate the structural degeneration of the bioprosthetic heart valves (BHVs). Using this approach, the critical length of the collagen fibrils l c and the interface shear acting on the fibrils in both BHV and natural aortic valve (AV) tissues under physiological loading conditions are calculated and presented. It is shown that the required critical fibril length to provide effective reinforcement to the natural AV and the BHV tissue is l c = 25.36 µm and l c = 66.81 µm, respectively. Furthermore, the magnitude of the required shear force acting on fibril interface to break a cross-linked fibril in the BHV tissue is shown to be 38 µN, while the required interfacial force to break the bonds between the fibril and the surrounding extracellular matrix is 31 µN. Direct correlations are underpinned between these values and the ultimate failure strength and the failure mode of the BHV tissue compared with the natural AV, and are verified against the existing experimental data. The analyses presented in this paper explain the role of fibril interface shear and critical length in regulating the biomechanics of the structural failure of the BHVs, for the first time. This insight facilitates further understanding into the underlying causes of the structural degeneration of the BHVs in vivo. |
Keywords | 0903 Biomedical Engineering; 0912 Materials Engineering; Biomedical Engineering |
Year | 2016 |
Journal | Journal of Materials Science: Materials in Medicine |
Journal citation | 27 (2) |
Publisher | Springer |
ISSN | 0957-4530 |
Digital Object Identifier (DOI) | https://doi.org/10.1007/s10856-015-5657-2 |
Publication dates | |
01 Feb 2016 | |
Publication process dates | |
Deposited | 21 Aug 2018 |
Accepted | 29 Dec 2015 |
Accepted author manuscript | License File Access Level Open |
Additional information | This is the accepted author manuscript for an article published by Springer Verlag in the Journal of Materials Science: Materials in Medicine. The full version is available online at: https://link.springer.com/article/10.1007/s10856-015-5657-2 |
https://openresearch.lsbu.ac.uk/item/87513
Download files
Accepted author manuscript
Evaluation_of_bioprosthetic_heart_valve_failure_using_a_matrix_fibril_shear_stress_transfer_approach.pdf | ||
License: CC BY 4.0 | ||
File access level: Open |
90
total views208
total downloads0
views this month1
downloads this month